Abbott Medical recalls Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the pati…
- Recall date
- October 1, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0551-2025
- FDA classification
- Class III
- Brand / firm
- Abbott Medical
- Sold / distributed
- Worldwide - U.S. Nationwide distribution in the states of ALABAMA, ALASKA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, ILLINOIS, INDIANA, KENTUCKY, MICHIGAN, MISSISSIPPI, MISSOURI, MONTANA, NEBRASKA, NEW JERSEY, NEW YORK, NORTH CAROLINA, OHIO, OREGON, PENNSY…
Why it was recalled
Due to heart transmitter not being able to upgrade to the current software version in the field after experiencing a power interruption during a prior over-the-wire software download.
Class III means the product violates regulations but is not likely to cause harm.
What was recalled
Merlin@home Transmitter REF EX1100 The Merlin@home" (M@h) transmitter is intended to transfer stored data from the patient s implanted heart device to Merlin.net" Patient Care Network (PCN) to facilitate remote monitoring by using inductive (EX1100 Model) or RF telemetry (EX1150 Model) to read device data and transmitting to the Merlin.net PCN server.
Get recall alerts
Free email alert whenever Abbott Medical has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Abbott Medical