Medical device recalls Moderate risk

Abbott Medical recalls The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming sys…

Recall date
November 21, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0753-2025
FDA classification
Class II
Brand / firm
Abbott Medical
Sold / distributed
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN TX, UT, VA, WA, WI, and WV. The countries of Austria, Belgium, Canada, Denma…

Why it was recalled

Due to a programmer software anomaly, during finalizing of the leadless pacemaker system, a specific sequence of programmer actions combined with a loss of telemetry during a small window (<2 seconds) may cause the finalization step to fail. If this occurs, the Programmer displays a loss of telemetry detected window, and there is no pacing.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

The Merlin PCS 3650 programmer Model 3330 The Merlin PCS programmer model 3650 is a portable, dedicated programming system which operates using the Merlin PCS Model 3330 software and is designed to interrogate, program, display data from, and test Abbott Medical implantable devices and leads.