Medical device recalls Moderate risk

ASP Global, LLC. dba Anatomy Supply Partners, LLC. recalls SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 ¿L; Prepared with Balanced Heparin, Model/Catalog Number: 06 0187 Pr…

Recall date
September 26, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0558-2026
FDA classification
Class II
Brand / firm
ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Sold / distributed
Distribution US nationwide and Canada.

Why it was recalled

Due to products manufactured under an inadequate quality system to include unvalidated manufacturing processes.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

SAFE-T-FILL Plastic Blood Gas Capillary Tubes, 230 ¿L; Prepared with Balanced Heparin, Model/Catalog Number: 06 0187 Product Description: A clear, plastic capillary blood collection tube