Medical device recalls High risk

B Braun Medical Inc recalls B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096

Recall date
March 19, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1799-2026
FDA classification
Class I
Brand / firm
B Braun Medical Inc
Sold / distributed
Worldwide distribution - US Nationwide and the country of Canada.

Why it was recalled

Potential for the accumulation of small air bubbles in the arterial line due to adherence of blood gases to the tubing under negative pressure.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

B. Braun Hemodialysis Bloodlines, STREAMLINE BLOODLINE SET FOR DIALOG. Model Number: SL-2010M2096