Baxter Healthcare Corporation recalls Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 016…
- Recall date
- September 17, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0233-2026
- FDA classification
- Class II
- Brand / firm
- Baxter Healthcare Corporation
- Sold / distributed
- Distribution US nationwide, Canada and Colombia.
Why it was recalled
Affected oral/axillary probes were inadvertently programmed with the rectal probe configuration. Affected probes will potentially display a lower temperature than the true value, which may lead to a delay in managing fever-related symptoms or infection.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Oral Probe (Product code 02893-000), accessory used with Welch Allyn SureTemp Plus 690 (Product codes 01690-200 and 01690-400) Thermometers and Welch Allyn SureTemp Plus 692 (01692-200) thermometers
Get recall alerts
Free email alert whenever Baxter Healthcare Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Baxter Healthcare Corporation