Medical device recalls Moderate risk

Beckman Coulter, Inc. recalls Access 2 Immunoassay Analyzer, Catalog Number 81600N

Recall date
February 5, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1386-2025
FDA classification
Class II
Brand / firm
Beckman Coulter, Inc.
Sold / distributed
Worldwide - US Nationwide distribution.

Why it was recalled

Beckman Coulter identified that an issue for Access 2 Sample Motors (B48016) due to improper alignment, causing the motors to rub against the sample wheelbase plate and index sensor. The issue was traced to a worn tool used in production (at the supplier), affecting Access 2 sample motors. This could lead to a potential delay in reporting patient test results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Access 2 Immunoassay Analyzer, Catalog Number 81600N