Medical device recalls Moderate risk

Beckman Coulter, Inc. recalls Access Testosterone assay, Catalog Number 33560

Recall date
May 5, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1901-2025
FDA classification
Class II
Brand / firm
Beckman Coulter, Inc.
Sold / distributed
US State: FL. UAE, Vietnam

Why it was recalled

Inadequate preventative maintenance (PM) performed on the scales of the RPL5 fill line, led to splashing (cross-contamination) between the wells of testosterone reagent packs lot 538081, which could potentially result in up to a 34% increase or a 23% decrease in patient results.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Access Testosterone assay, Catalog Number 33560