Beckman Coulter Inc. recalls DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The…
- Recall date
- February 26, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1517-2025
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter Inc.
- Sold / distributed
- Worldwide Distribution: U.S. (nationwide) to states of: AL, CA, CO, GA, ID, IL, IN, LA, MA, AI, MN, MS, MT, NC, NM, NV, NY, OK, OR, PA, PR, SC, TN, TX, WA, and WI; O.U.S. (foreign) to countries of: Australia, Belgium, Bosnia/Herzegovina, Canada, Croatia, Egypt, Ethiopia, France, Germany, Ghana, Ind…
Why it was recalled
Beckman Coulter has determined that device software versions V1.3, V1.4, V1.4.1, V1.4.2, V1.4.3. have a defect due of which, when all constituent tests of a calculated result are rerun, either manually ordered by the operator at the analyzer or automatically triggered by predefined rules in System Configuration, the calculated result will not be recalculated using the constituent tests rerun results. The system will report only the calculated result using the initial constituent tests results to the operator and LIS/Remisol. The issue was identified by Beckman Coulter internal engineering testing. The defect may cause a delay in reporting patient results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DxC 500 AU Clinical Chemistry Analyzer, Catalog Numbers/UDI codes: C63519 / 14987666545058 C63520 / 14987666545065 The Beckman Coulter DxC 500 AU Clinical Chemistry Analyzer is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Get recall alerts
Free email alert whenever Beckman Coulter Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Beckman Coulter Inc.