Beckman Coulter, Inc. recalls DxI 600 Access Immunoassay Analyzer W/Dual Gantry, Part Number A71461
- Recall date
- June 25, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2270-2025
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter, Inc.
- Sold / distributed
- Worldwide distribution - US Nationwide.
Why it was recalled
Beckman Coulter has identified that the UniCel DxI 600/800 systems with linear slide PnP gantries running software versions 5.7 and up faced incorrect PnP X-motor current settings where the PnP X-motor current incorrectly resets to 1.5A (instead of the required 2.0A) because the software overwrites the factory setting during installs or upgrades. This anomaly causes increased PnP motion errors which could lead to delays in reporting patient sample results if unnoticed.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.