Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278
- Recall date
- February 6, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1387-2025
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter, Inc.
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of FL, TX, WA.
Why it was recalled
Beckman Coulter identified that an issue for DxI 9000 tip stocker interface (TSI) and reagent storage interface (RSI) boards with the firmware version 1.1.0.0 can cause sub-system communication failures. This failure could result in a delay in reporting patient results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DxI 9000 Access Immunoassay Analyzer, Catalog Number C11137, with Component PCB, TSI/RSI firmware C29278