Beckman Coulter, Inc. recalls DxI 9000 Access Immunoassay Analyzer, Part Number C11137

Recall date
September 19, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0216-2026
FDA classification
Class II
Brand / firm
Beckman Coulter, Inc.
Sold / distributed
Worldwide - US Nationwide distribution.

Why it was recalled

Beckman Coulter has confirmed a software anomaly on DxI 9000 Access Immunoassay Analyzer and the DxC 500i Clinical Analyzer that may prevent correct application of user-specified manual dilution factors when ordering additional tests for samples already "In Progress" using the same "Sample ID number". This can lead to false low or high patient results and potential delays in reporting, risking erroneous diagnoses if undetected.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

DxI 9000 Access Immunoassay Analyzer, Part Number C11137