Medical device recalls Moderate risk

Beckman Coulter Inc. recalls UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

Recall date
August 13, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2574-2025
FDA classification
Class II
Brand / firm
Beckman Coulter Inc.
Sold / distributed
Worldwide - US Nationwide distribution including in the states of AL, AZ, CA, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NJ, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, WA, WI, WV and the countries of Algeria, Argentina, Australia, Bahrain, Bosnia and Herzegovina, B…

Why it was recalled

Hematology analyzers configured with HGB photometers may result in erroneously high hemoglobin (HGB) results on samples with elevated white blood cell (WBC) counts.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

UniCel DxH 800 COULTER Cellular Analysis System, configured with the Led HGB Photometer REFs:B24802, 629029, B63322

Get recall alerts