Beckman Coulter Mishima K.K. recalls DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples…
- Recall date
- July 10, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2691-2024
- FDA classification
- Class II
- Brand / firm
- Beckman Coulter Mishima K.K.
- Sold / distributed
- Worldwide - U.S. Nationwide distribution in the states of AL, CA, GA, HI, IL, MI, MN, MS, NE, NY, OR, PA and SC. The countries of Lebanon, South Africa.
Why it was recalled
Due to a software issues, after the instrument processes 250 racks cumulatively, any subsequent racks with samples requiring rerun/reflex will be held in the Sample Handler's Buffer area and and error code "9000" will be reported. This issue may cause a delay of results.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
DxC 500 AU Clinical Chemistry Analyzer, REF C63520 is an automated chemistry analyzer that measures analytes in samples, in combination with appropriate reagents, calibrators, quality control (QC) material and other accessories. This system is for in vitro diagnostic use only.
Get recall alerts
Free email alert whenever Beckman Coulter Mishima K.K. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Beckman Coulter Mishima K.K.