Medical device recalls Moderate risk

Becton, Dickinson and Company, BD Bio Sciences recalls BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD

Recall date
July 8, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2237-2025
FDA classification
Class II
Brand / firm
Becton, Dickinson and Company, BD Bio Sciences
Sold / distributed
Worldwide - US Nationwide distribution in the states of Tennessee, New York, Massachusetts and the countries of Argentina, Belgium, Botswana, Brazil, Canada, France, Germany, Lithuania, Peru, Philippines, United Kingdom.

Why it was recalled

Control kits contain incorrect standard deviation (SD) values for the low and high control beads which compromises the accurate calculation of bead counts.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Trucount Controls, 30 Test - REF: 664343, RxOnly, IVD