Becton Dickinson & Co. recalls BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantita…
- Recall date
- November 6, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0621-2025
- FDA classification
- Class II
- Brand / firm
- Becton Dickinson & Co.
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of CA, GA, IL, KY, MA, MD, NE, PA and the countries of Canada, Belgium, Singapore, Malaysia.
Why it was recalled
BD identified through potency testing as part of a stability test request to monitor Ampicillin AM-2 due to decrease in potency results of 65% at 18 months, may result in falsely resistant result for ampicillin susceptibility
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) - In-Vitro BD BBL Sensi Disc Ampicillin 2 ¿g (AM-2) are used for semi-quantitative in vitro susceptibility testing by the agar disc diffusion test procedure of common, rapidly growing and certain fastidious bacterial pathogens. Catalog Number: 231263
Get recall alerts
Free email alert whenever Becton Dickinson & Co. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Becton Dickinson & Co.