Medical device recalls Moderate risk

Belmont Instrument LLC recalls 3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoir…

Recall date
March 31, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1678-2025
FDA classification
Class II
Brand / firm
Belmont Instrument LLC
Sold / distributed
Nationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI. International distribution to Czechia, Italy, Norway.

Why it was recalled

Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.