Medical device recalls High risk

Boston Scientific Corporation recalls ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Nu…

Recall date
December 12, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0822-2025
FDA classification
Class I
Brand / firm
Boston Scientific Corporation
Sold / distributed
US (Nationwide) including Puerto Rico and US Virgin Islands and OUS (foreign) to countries of: Guam and Northern Mariana Islands.

Why it was recalled

A subset of devices from the ACCOLADE Family of Pacemakers and CRT-Ps, built before September 2018, have an increased potential to initiate Safety Mode during telemetry or other normal, higher-power operations due to latent high battery impedance.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

ACCOLADE DR Pacemaker, SL (Model Number L301), SL MRI (Model Number L311), EL (Model Number L321), and EL MRI (Model Number L331)

Get recall alerts

Free email alert whenever Boston Scientific Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Boston Scientific Corporation