Boston Scientific Corporation recalls Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H7…
- Recall date
- December 3, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1042-2026
- FDA classification
- Class II
- Brand / firm
- Boston Scientific Corporation
- Sold / distributed
- US Nationwide distribution in the states of MA, WV, OH, CA, MN, NY, VA, IL TX, FL, IL.
Why it was recalled
Boston Scientific is initiating the removal of certain batches of Coyote" Over-the-Wire (OTW) Percutaneous Transluminal Angioplasty (PTA) Balloon Dilatation Catheter devices due to the potential for inflation/deflation difficulty during use in the peripheral vasculature. An internal investigation found that a manufacturing process variation resulted in catheters that may not properly inflate/deflate. This was limited to Coyote OTW batches manufactured from 12Sep2025 through 13Oct2025. The processing variation has since been addressed. No other material numbers or batches are impacted by this removal.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Boston Scientific Coyote OVER-THE-WIRE PTA Balloon Dilation Catheter, 4F, 1) 2.5mm x120mm, 150 cm, UPM/Model No. REF H74939186251210; 2) 3.5mm x120mm, 150 cm, UPM/Model No. REF H74939186351210; Catheter, Angioplasty, Peripheral, Transluminal
Get recall alerts
Free email alert whenever Boston Scientific Corporation has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Boston Scientific Corporation