Medical device recalls High risk

Boston Scientific Corporation recalls *HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;

Recall date
December 19, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1163-2026
FDA classification
Class I
Brand / firm
Boston Scientific Corporation
Sold / distributed
Worldwide distribution. US Nationwide including Puerto Rico; Algeria, Argentina, Austria, Belgium, Brazil, Canada, CHILE, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, JAPAN, Jordan, Kuwait, Lebanon…

Why it was recalled

Increased reports of stent deployment and expansion issues. The most common outcome reasonably foreseeable resulting from difficulty deploying the stent is prolongation of the procedure to exchange the device for a new one. The most serious reasonably foreseeable outcome in cases where the first flange is unable to deploy or expand would be a cascade of events resulting in additional intervention, either endoscopic or surgical, to remove the stent and close the puncture site.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

*HOT AXIOS Stent and Electrocautery-Enhanced Delivery System 20mm x 10mm; Material Number (UPN): M00553560;

Get recall alerts

Free email alert whenever Boston Scientific Corporation has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Boston Scientific Corporation