Boston Scientific Corporation recalls WATCHMAN TruSeal Access System ANT, US, Material Number (UPN) M635TU70040; intended to provide vascular and transseptal…
- Recall date
- July 29, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2433-2025
- FDA classification
- Class I
- Brand / firm
- Boston Scientific Corporation
- Sold / distributed
- Worldwide.
Why it was recalled
Boston Scientific has identified a higher likelihood of air embolism events when procedures are performed without positive pressure-controlled ventilation. According to published literature and clinical data, in percutaneous procedures requiring transseptal access to the left atrium when conscious or deep sedation is used, patients have an approximately three-times higher risk (U.S. study) of negative left atrium pressure and air ingress. This risk is especially prevalent in patients with pre-existing low left atrial pressure, hypovolemia, and partial upper airway collapse.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
WATCHMAN TruSeal Access System ANT, US, Material Number (UPN) M635TU70040; intended to provide vascular and transseptal access for all WATCHMAN Left Atrial Appendage Closure Devices with Delivery Systems.