Medical device recalls Moderate risk

Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T Model MRT-3020/MEXL-3020

Recall date
December 19, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1017-2026
FDA classification
Class II
Brand / firm
Canon Medical System, USA, INC.
Sold / distributed
U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.

Why it was recalled

There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

VANTAGE GALAN 3T Model MRT-3020/MEXL-3020