Canon Medical System, USA, INC. recalls VANTAGE GALAN 3T Model MRT-3020/MEXL-3020
- Recall date
- December 19, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1017-2026
- FDA classification
- Class II
- Brand / firm
- Canon Medical System, USA, INC.
- Sold / distributed
- U.S. Nationwide distribution in the states of CA, CO, FL, IA, IL, IN, KS, MD, MO, MT, NV, NY, OH, OR, PA, PR, TX, VA, and WI.
Why it was recalled
There is a potential for formation of ice occurring in the venting system of the superconducting magnet of the MRI systems. if a quench occurs in this condition, helium gas may not be able to escape via the designed vent paths, and the pressure in the helium vessel may rise. This pressure build-up could eventually rupture the helium vessel, causing helium gas to be released into the MR scanning room.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
VANTAGE GALAN 3T Model MRT-3020/MEXL-3020