Medical device recalls Moderate risk

Cardinal Health 200, LLC recalls Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" S…

Recall date
September 18, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0206-2026
FDA classification
Class II
Brand / firm
Cardinal Health 200, LLC
Sold / distributed
Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of A…

Why it was recalled

Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Salem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" SALEM SUMP" PVC TUBE 12FR 48IN 8888264945E ENFit" SALEM SUMP" PVC TUBE 14FR 48IN 8888264960E ENFit" Salem Sump" PVC Tube 16Fr 48IN 8888264986E ENFit" SALEM SUMP" PVC TUBE 18FR 48IN 8888266106E PVC SALEM SUMP WITH ENFIT & ARV 10FR 36IN 8888266114E PVC SALEM SUMP WITH ENFIT & ARV 12FR 48IN 8888266122E PVC SALEM SUMP WITH ENFIT & ARV 14FR 48IN 8888266130E PVC SALEM SUMP WITH ENFIT & ARV 16FR 48IN 8888266148E PVC SALEM SUMP WITH ENFIT & ARV 18FR 48IN 8888268060E ENFit" SALEM SUMP" PVC TUBE 6FR 24IN 8888268086E ENFit" Salem Sump" PVC Tube 8Fr 24IN