Medical device recalls Moderate risk

CareFusion 303, Inc. recalls BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a mod…

Recall date
November 6, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1217-2026
FDA classification
Class II
Brand / firm
CareFusion 303, Inc.
Sold / distributed
U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.

Why it was recalled

Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.

Get recall alerts

Free email alert whenever CareFusion 303, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: CareFusion 303, Inc.