CareFusion 303, Inc. recalls BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a mod…
- Recall date
- November 6, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1217-2026
- FDA classification
- Class II
- Brand / firm
- CareFusion 303, Inc.
- Sold / distributed
- U.S. Nationwide distribution in the states of AZ, IN, OH, and VA.
Why it was recalled
Due to a number of modules flashed with a date and time associated with the daylight savings time (DST) adjustment that may cause connectivity issue with hospital networks. This issue only affects the Interoperability workflow for devices with a DST invalid timestamp, impacting connectivity of the PCU to the hospital network
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BD Alaris" Pump Module model 8100 UDI-DI code: 10885403810046 The BD AlarisTM System with Guardrails" Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration: intravenous (IV), intra-arterial (IA), subcutaneous, epidural, or irrigation of fluid spaces. The BD AlarisTM System includes the PC Unit (PCU) and one or more of the following: Pump Module, Syringe Module, end-tidal CO2 (EtCO2) Module, Auto-ID Module, patient-controlled analgesia (PCA) Module, and associated software applications.
Get recall alerts
Free email alert whenever CareFusion 303, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: CareFusion 303, Inc.