Medical device recalls Moderate risk

CareFusion 303, Inc. recalls BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring…

Recall date
January 23, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1231-2025
FDA classification
Class II
Brand / firm
CareFusion 303, Inc.
Sold / distributed
U.S.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and Guantanamo Bay, Cuba O.U.S.: Canada

Why it was recalled

Due to User Manual/Direction for Use incorrectly instructing user to leave the safety clamp in the open position when reloading an infusion set. This issue potentially could result in an accidental over infusion if the roller clamp is left open.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

BD Alaris Pump Module REF 8100 The BD Alaris System with Guardrails Suite MX is a modular infusion pump and monitoring system for the continuous or intermittent administration of fluids to adult, pediatric, and neonatal patients through clinically accepted routes of administration

Get recall alerts

Free email alert whenever CareFusion 303, Inc. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: CareFusion 303, Inc.