Civco Medical Instruments Co. Inc. recalls BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagn…
- Recall date
- May 1, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1892-2025
- FDA classification
- Class II
- Brand / firm
- Civco Medical Instruments Co. Inc.
- Sold / distributed
- OUS Distribution to countries of: Japan, Taiwan, France, Georgia, Germany and Denmark
Why it was recalled
During the attachment of the needle guide to the covered probe, the tight tolerance and fit of the needle guide can cause a pinhole breach in the transducer cover when latching the guide in place.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
BX2 Needle Guide, Part Number 644-094, tool for performing needle-guided (or catheter) procedures with the use of diagnostic ultrasound transducers