HSG Procedure Tray recalled over sterility concerns
- Recall date
- June 11, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Official notice title
- CooperSurgical, Inc. recalls HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterograp…
- Recall number
- Z-2135-2025
- FDA classification
- Class II
- Brand / firm
- CooperSurgical, Inc.
- Sold / distributed
- US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington DC.
Why it was recalled
The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterography procedures.