Medical device recalls Moderate risk

HSG Procedure Tray recalled over sterility concerns

Recall date
June 11, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Official notice title
CooperSurgical, Inc. recalls HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterograp…
Recall number
Z-2135-2025
FDA classification
Class II
Brand / firm
CooperSurgical, Inc.
Sold / distributed
US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, HI, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and Washington DC.

Why it was recalled

The HSG Procedure Trays and Endosee System Convenience Kits contain non-sterile iodine pouches, however the peel away lid labels incorrectly identify the iodine pouches as sterile.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

HSG Procedure Tray. Model Number: 6050T. The HSG Procedure Tray is for use in hysterosalpingography and sonohysterography procedures.