DATEX--OHMEDA, INC. recalls GE Heathcare Giraffe OmniBed; neo natal incubator
- Recall date
- September 3, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0047-2025
- FDA classification
- Class I
- Brand / firm
- DATEX--OHMEDA, INC.
- Sold / distributed
- Worldwide distribution.
Why it was recalled
GE HealthCare has become aware that for certain Giraffe OmniBed and Giraffe OmniBed Carestation devices, the screw that secures the doors that cover the warmer heaters may not have been torqued to specification. This may lead to the doors becoming loose.
Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.
What was recalled
GE Heathcare Giraffe OmniBed; neo natal incubator
Get recall alerts