Medical device recalls Moderate risk

Defibtech, LLC recalls Defibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indi…

Recall date
October 14, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0502-2025
FDA classification
Class II
Brand / firm
Defibtech, LLC
Sold / distributed
Ireland

Why it was recalled

Within the Irish Market, a Field Safety Notice is being issued due to a select number of authorized labelled Defibtech AED Pads, which have had the authorized label removed and an unauthorized label applied, could lead to expired Defibtech AED Pads appearing as though they are within the acceptable shelf life. There is a possibility of hydrogel drying, which may result in poor electrical connection to the patient, which can reduce the energy delivered to the patient.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Defibtech DDP-100 Adult Defibrillation Pads For use with AED Models: Lifeline and Lifeline Auto (DDU-1:XX Series)- Indicated for use on victims of sudden cardiac arrest (SCA)

Get recall alerts

Free email alert whenever Defibtech, LLC has a new recall — straight from official government data. Unsubscribe anytime.