Diagnostica Stago, Inc. recalls STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of t…
- Recall date
- July 21, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2476-2025
- FDA classification
- Class II
- Brand / firm
- Diagnostica Stago, Inc.
- Sold / distributed
- Worldwide - US Nationwide distribution in the states of AL, AZ, CA, CO, FL, GA, LA, MD, ME, MN, MO, MT, NC, OH, OK, OR, PA, PR, SD, VT, WI, WY and the countries of Algeria, Argentina, Austria, Brazil, Bulgaria, Canada, Chile, Colombia, Costa Rica, France, French Polynesie, Guadeloupe, India, Iran,…
Why it was recalled
Potential for contamination of STA-LIQUID ANTI-Xa by Stago Fibrinogen reagents in UFH and LMWH applications.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
STA - Fibrinogen 5. Model/Catalog Number: 00674. The STA - Fibrinogen 5 kit is intended for use with the analyzers of the STA-R, STA Compact and STA Satellite families for the quantitative determination of fibrinogen levels in human citrated plasma by the clotting method of Clauss.