Medical device recalls High risk

Edwards Lifesciences, LLC recalls Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Ver…

Recall date
May 14, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2080-2025
FDA classification
Class I
Brand / firm
Edwards Lifesciences, LLC
Sold / distributed
Worldwide - US Nationwide distribution including in the states of ALABAMA, ARIZONA, ARKANSAS, CALIFORNIA, COLORADO, CONNECTICUT, DELAWARE, DISTRICT OF COLUMBIA, FLORIDA, GEORGIA, HAWAII, IDAHO, ILLINOIS, INDIANA, IOWA, KANSAS, KENTUCKY, LOUISIANA, MAINE, MARYLAND, MASSACHUSETTS, MICHIGAN, MINNESOTA…

Why it was recalled

due to complaints received for 3mm to 4mm section wire exposed/protruding from the outside of the cannula body.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Brand Name: Optisite Arterial Cannula Product Name: Optisite Arterial Cannula Model/Catalog Number: OPTI16 Software Version: N/A Product Description: Cardiac cannula. The Edwards OptiSite Arterial Perfusion Cannluae provides or establishes arterial perfusion in the extracorporeal circuit for d 6 hours. Component: No.