Erbe Medical, LLC recalls ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
- Recall date
- March 31, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-2064-2026
- FDA classification
- Class II
- Brand / firm
- Erbe Medical, LLC
- Sold / distributed
- US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.
Why it was recalled
Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
ERBEFLO 2, Endoscopy Pump Tubing/Cap Set, REF: 20325-222
Get recall alerts
Free email alert whenever Erbe Medical, LLC has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Erbe Medical, LLC