Medical device recalls High risk

Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environm…

Recall date
January 10, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1019-2025
FDA classification
Class I
Brand / firm
Fresenius Kabi USA, LLC
Sold / distributed
US distribution to CA, CO, GA, ID, MI, MS, NE, NJ, NV, OK, TX, UT, VA, WA, WI.

Why it was recalled

Large Volume Pump Software, version 5.9.2 and earlier has potential for the following anomalies: 1. The pump may become nonfunctional if during an alarm condition the Pause Audio option is repeated 70 time or more. 2. If a secondary infusion is started at the exact moment a primary infusion completes and VTBI reaches 0, it will switch to primary. The primary infusion will infuse at the previously programed primary rate and continue until the infusion is stopped or the bag is empty.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Ivenix Infusion System (IIS), LVP Software LVP-SW-0005. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.

Get recall alerts

Free email alert whenever Fresenius Kabi USA, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Fresenius Kabi USA, LLC