Fujirebio Diagnostics, Inc. recalls Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applic…
- Recall date
- December 11, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1306-2026
- FDA classification
- Class II
- Brand / firm
- Fujirebio Diagnostics, Inc.
- Sold / distributed
- US Nationwide distribution in the states of Arizona, California, Florida, Indiana, Kansas, Kentucky, Maryland, Massachusetts, Minnesota, New Jersey, Pennsylvania, Texas, Washington.
Why it was recalled
Customers may experience inaccurate test results by observing a higher frequency of classification in the indeterminate and positive zone and/or low specificity when compared back to FDA-cleared CSF ratio/PET imaging, due to falsely elevated positive ratio or falsely elevated indeterminate ratio results causing an incorrect classification of a patient as having amyloid pathology associated with Alzheimer s disease (positive result) and/or requiring more clinical testing of that patient (indeterminate result).
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Product Name: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators Model/Catalog Number: 235454 Software Version: Not Applicable Product Description: Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: The ¿-Amyloid 1-42 plasma concentration of a specimen is automatically calculated from the calibration curve, which is automatically calculated from calibration data. The result of the calculation is reported in pg/mL. Lumipulse G ¿-Amyloid 1-42-N Plasma Calibrators: Liquid (Frozen), 1 ¿ 1.5 mL (4 Concentrations) CAL 1 0 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 2 30 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 3 100 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator CAL 4 1000 pg/mL ¿-Amyloid 1-42-N Plasma Calibrator Contains Tris buffer with protein (bovine) and chemical stabilizers. Preservative: 0.1% ProClin 950. Component: Yes. Lumipulse G pTau217/B-Amyloid 1-42 Plasma Ratio
Get recall alerts
Free email alert whenever Fujirebio Diagnostics, Inc. has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Fujirebio Diagnostics, Inc.