GE Healthcare (China) Co., Ltd. recalls Optima MR360, NMRI system
- Recall date
- January 31, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1241-2025
- FDA classification
- Class II
- Brand / firm
- GE Healthcare (China) Co., Ltd.
- Sold / distributed
- Domestic US distribution nationwide. International distribution worldwide.
Why it was recalled
For certain MR systems, a scan can be resumed following a Power Monitor trip when the "6-minute average SAR" is above the limit indicated on the "SAR Display". The user interface may incorrectly enable a scan to proceed earlier than intended. If resumed by the operator, such a scan can result in potential excessive tissue heating, especially if the patient is in contact with a conductive material or has an implant.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Optima MR360, NMRI system