Medical device recalls High risk

GE Medical Systems China Co., Ltd. recalls GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.

Recall date
March 21, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1632-2025
FDA classification
Class I
Brand / firm
GE Medical Systems China Co., Ltd.
Sold / distributed
Worldwide distribution. US nationwide including Puerto Rico, Albania, Algeria, Argentina, Armenia, Australia, Austria, Bahrain, Bangladesh, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Cambodia, Canada, Chile, China, Colombia, Costa Rica, Cote d'Ivoire, Croatia, Cyprus, Czech Republi…

Why it was recalled

Certain Carestation 620/650/650c and 750/750c Anesthesia Delivery Systems will not provide effective ventilation in Volume Control Ventilation (VCV) mode. In these systems, effective ventilation can be achieved in Pressure Control Ventilation (PCV) or Pressure Control Ventilation Volume Guarantee (PCV-VG) modes or with Manual ventilation.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

GE Healthcare CARESTATION 650C A1, Model/REF Number 1012-9655-000. Anesthesia system.

Get recall alerts

Free email alert whenever GE Medical Systems China Co., Ltd. has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Medical Systems China Co., Ltd.