Medical device recalls Moderate risk

GE Medical Systems Information Technologies Inc recalls GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Teleme…

Recall date
April 24, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2244-2026
FDA classification
Class II
Brand / firm
GE Medical Systems Information Technologies Inc
Sold / distributed
Worldwide distribution.

Why it was recalled

GE HealthCare has become aware of an unlikely situation where potential loss of ECG and SpO2 monitoring involving ApexPro CARESCAPE Telemetry Server (CTS) v5.0 and v6.0, can result from a persistent NO COMM or OFF NETWORK condition at the CARESCAPE Central Station or CIC Pro Clinical Information Center (Central Station).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GE HealthCare IOU ASSY ARK2250 CARESCAPE TELEMETRY SERVER, REF #2108729-901, Product Code SS9; ApexPro CARESCAPE Telemetry Server (CTS) v5.0; Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)

Get recall alerts

Free email alert whenever GE Medical Systems Information Technologies Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Medical Systems Information Technologies Inc