Medical device recalls Moderate risk

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging

Recall date
September 12, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0293-2026
FDA classification
Class II
Brand / firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Sold / distributed
Worldwide

Why it was recalled

GE HealthCare has become aware that certain Nuclear Medicine systems that are past the End of Guaranteed Service could have been transported or relocated without adequate detector support. This could result in excessive stress being applied to the detector mounting mechanisms, compromising their integrity. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Elscint SPX6; dual-head nuclear medicine gamma camera designed specifically for cardiac imaging