Medical device recalls Moderate risk

GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING recalls GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission

Recall date
December 18, 2024
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-0896-2025
FDA classification
Class II
Brand / firm
GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING
Sold / distributed
Worldwide

Why it was recalled

GE HealthCare has become aware that certain Infinia, VariCam/VG/VH, Brivo NM 615, Discovery NM 630, Discovery NM/CT 670, and Optima NM/CT 640 Nuclear Medicine systems (see Affected Products Details and Appendices) could have been transported or relocated without proper detector support, which can result in excessive load being applied to the detector mounting mechanisms. If this occurs, it can potentially result in a detector fall and lead to life-threatening bodily injury.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GE Healthcare Optima NM/CT 640, Model/Catalog Numbers: 1) H3100JZ; System, Tomography, Computed, Emission

Get recall alerts

Free email alert whenever GE MEDICAL SYSTEMS ISRAEL, FUNCTIONAL IMAGING has a new recall — straight from official government data. Unsubscribe anytime.