Medical device recalls Moderate risk

GE Medical Systems, LLC recalls Discovery MR750w 3.0T

Recall date
May 23, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2132-2025
FDA classification
Class II
Brand / firm
GE Medical Systems, LLC
Sold / distributed
Worldwide distribution - US Nationwide and the country of China.

Why it was recalled

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Discovery MR750w 3.0T

Get recall alerts

Free email alert whenever GE Medical Systems, LLC has a new recall — straight from official government data. Unsubscribe anytime.