Medical device recalls Moderate risk

GE Medical Systems, LLC recalls GE HealthCare OEC 9800.

Recall date
July 2, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-2341-2025
FDA classification
Class II
Brand / firm
GE Medical Systems, LLC
Sold / distributed
US Nationwide distribution.

Why it was recalled

GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

GE HealthCare OEC 9800.