Medical device recalls Moderate risk

GE Medical Systems, LLC recalls Optima XR646 HD. X-Ray imaging system.

Recall date
January 28, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1046-2025
FDA classification
Class II
Brand / firm
GE Medical Systems, LLC
Sold / distributed
Worldwide Distribution: US (nationwide) including Puerto Rico and OUS (foreign) countries to: Canada, Chile, China, Colombia, Coast Rica, Croatia, Cyprus, Czech Republic, Czechia, Dominican Republic, Egypt, Estonia, Finland, France, Germany, Gibraltar, Greece, Honduras, Hong Kong, Hungary, Iceland,…

Why it was recalled

The user can continue to take the next exposure after reaching the Automatic Exposure Control (AEC) limit, without taking any further actions to acknowledge or adjust exposure techniques. In all cases, a message still appears on the screen when the AEC limit is reached.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Optima XR646 HD. X-Ray imaging system.

Get recall alerts

Free email alert whenever GE Medical Systems, LLC has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: GE Medical Systems, LLC