Getinge Usa Sales Inc recalls MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during…
- Recall date
- October 3, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0463-2025
- FDA classification
- Class II
- Brand / firm
- Getinge Usa Sales Inc
- Sold / distributed
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bhutan, Brazil, Canada, Chile, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, France, Germany, Greece, India, Israel, Japan, Jordan, Korea, Republic Of Kuwait, "Macedonia, The Former Yugoslav Republic Of…
Why it was recalled
Error code "50037" occurs on the IR-Hand Control when a MEERA-table is controlled, and the table stops the movement. This error may occur sporadically when the MEERA-table is controlled via an IR-Hand Control and may result in procedural delay
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
MEERA CL mobile operating table - Designed for the placement and positioning of the patient immediately before, during and after surgical interventions, as well as for examination and treatment Item Number: 700001F0
Get recall alerts
Free email alert whenever Getinge Usa Sales Inc has a new recall — straight from official government data. Unsubscribe anytime.
More recalls from: Getinge Usa Sales Inc