Medical device recalls High risk

Getinge Usa Sales Inc recalls Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Fl…

Recall date
January 15, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1074-2025
FDA classification
Class I
Brand / firm
Getinge Usa Sales Inc
Sold / distributed
Domestic distribution to Kentucky and Connecticut.

Why it was recalled

Sevoflurane used with the vaporizer may degrade to hydrogen fluoride, which could present a risk to patients or health care providers through inhalation and/or skin exposure. Potential harms include irritation of respiratory tract, which in worst case may lead to lung edema and/or severe hypocalcemia, blistering, superficial ulceration, and/or hypomagnesemia.

Class I is the FDA’s most serious recall level: there is a reasonable probability that using this product will cause serious harm or death.

What was recalled

Vaporizer Sevoflurane, Quik-Fil. Model Number: 6682285. Device is a component of Flow-i C20, Flow-i C30, Flow-i C40, Flow-c, Flow-e. For use during anesthesia.

Get recall alerts

Free email alert whenever Getinge Usa Sales Inc has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Getinge Usa Sales Inc