Medical device recalls Moderate risk

Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software Ve…

Recall date
December 9, 2025
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1080-2026
FDA classification
Class II
Brand / firm
Heraeus Medical GmbH (Dental Division)
Sold / distributed
International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.

Why it was recalled

increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Brand Name: PALACOS Product Name: PALACOS R pro 40; PALACOS R pro 80 Model/Catalog Number: 5081286; 5081287 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A

Get recall alerts

Free email alert whenever Heraeus Medical GmbH (Dental Division) has a new recall — straight from official government data. Unsubscribe anytime.

More recalls from: Heraeus Medical GmbH (Dental Division)