Heraeus Medical GmbH (Dental Division) recalls Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Softw…
- Recall date
- December 9, 2025
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1082-2026
- FDA classification
- Class II
- Brand / firm
- Heraeus Medical GmbH (Dental Division)
- Sold / distributed
- International distribution in the countries of Aland Islands, Australia, Belgium, Denmark, Finland, Germany, Ireland, Italy, Liechtenstein, Netherlands, New Zealand, Norway, Portugal, Spain, Sweden, Switzerland, United Kingdom.
Why it was recalled
increase of complaints regarding the breakage of the ampoules within their device systems. An extensive investigation revealed that multiple factors have an impact on the ampoule breakage. An unfortunate combination of these influence factors can lead to unbreakable ampoules. If none or not all ampoules break, the product cannot be used on the patient since no or no appropriate cement dough is formed.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Brand Name: PALACOS Product Name: PALACOS MV+G pro 40; PALACOS MV+G pro 80 Model/Catalog Number: 5081290; 5081291 Software Version: N/A Product Description: Radiopaque, poly(methyl methacrylate)-based (PMMA) bone cement, pre-filled into a mixing and application system, suitable for use with or without vacuum (ready to mix). Component: N/A