Medical device recalls Moderate risk

Hologic, Inc recalls Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: C…

Recall date
March 23, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1949-2026
FDA classification
Class II
Brand / firm
Hologic, Inc
Sold / distributed
Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.

Why it was recalled

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.