Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment; Part Number: 210490

Recall date
February 2, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1572-2026
FDA classification
Class II
Brand / firm
Howmedica Osteonics Corp.
Sold / distributed
Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;

Why it was recalled

A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

MICS3 Angled Sagittal Saw Attachment; Part Number: 210490