Howmedica Osteonics Corp. recalls MICS3 Angled Sagittal Saw Attachment; Part Number: 210490
- Recall date
- February 2, 2026
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-1572-2026
- FDA classification
- Class II
- Brand / firm
- Howmedica Osteonics Corp.
- Sold / distributed
- Domestic: Nationwide Distribution; International: Australia, Germany, Belgium, France, Spain UK, New Zealand;
Why it was recalled
A potential issue was identified with the torque strength on the external screws of the MICS3 Angled Sagittal Saw Attachment. These screws, which secure the cover to the Angled Saw Housing, may loosen during use and detach from the attachment.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
MICS3 Angled Sagittal Saw Attachment; Part Number: 210490