Howmedica Osteonics Corp. recalls Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications du…
- Recall date
- September 4, 2024
- Source
- U.S. Food & Drug Administration (FDA) — enforcement report
- Recall number
- Z-0023-2025
- FDA classification
- Class II
- Brand / firm
- Howmedica Osteonics Corp.
- Sold / distributed
- US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Ur…
Why it was recalled
Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.
Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.
What was recalled
Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-200