Medical device recalls Moderate risk

I.T.S. GmbH recalls Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle…

Recall date
January 23, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1593-2026
FDA classification
Class II
Brand / firm
I.T.S. GmbH
Sold / distributed
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Why it was recalled

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

Claviculaplate with Angular Stability, with the below descriptions and associated article numbers. 1. Anterior Clavicle Plate, Medial; Article Numbers: 21116-8, 21117-10, 21118-10. 2. Clavicle Plate; Article Numbers: 21121-6, 21121-8. 3. Clavicle Plate, 3.5mm; Article Numbers: 21122-10, 21122-6, 21122-8. 4. Clavicle Plate, Lateral; Article Numbers: 21123-4, 21124-4, 21127-6, 21128-6.