Medical device recalls Moderate risk

I.T.S. GmbH recalls FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding…

Recall date
January 23, 2026
Source
U.S. Food & Drug Administration (FDA) — enforcement report
Recall number
Z-1605-2026
FDA classification
Class II
Brand / firm
I.T.S. GmbH
Sold / distributed
US Nationwide distribution in the states of AL, AZ, BVI, CA, CO, FL, GA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NE, NJ, NM, NV, OH, OK, PA, SC, TX, USVI, WV, W.

Why it was recalled

Updated MRI safety testing demonstrates higher RF-induced temperature increases under certain MRI conditions than previously reflected in the Instructions for Use (IFU).

Class II means using this product may cause temporary or medically reversible health problems; the chance of serious harm is remote.

What was recalled

FLS - Foot Locking Plates System and HOL - Hallux Osteotomy Locking Plate with the below descriptions and corresponding article numbers. 1. FLS - Foot Locking Plates System, Cancellous Stabilization Screw; Article Numbers: 37303-10, 37303-12, 37303-14, 37303-16, 37303-18, 37303-20, 37303-22, 37303-24, 37303-26, 37303-28, 37303-30. 2. FLS - Foot Locking Plates System, Cortical Screw; Article Numbers: 32271-10, 32271-12, 32271-14, 32271-16, 32271-18, 32271-20, 32271-22, 32271-24, 32271-26, 32271-28, 32271-32, 32271-34. 3. FLS - Foot Locking Plates System, Cortical Stabilization Screw; Article Numbers: 37304-10, 37304-12, 37304-14, 37304-16, 37304-18, 37304-20, 37304-22, 37304-24. 4. FLS - Foot Locking Plates System, FLS Plate, Curved; Article Numbers: 21413-6. 5. FLS - Foot Locking Plates System, FLS Plate, H-Shape; Article Numbers: 21431-ME. 6. FLS - Foot Locking Plates System, FLS Plate, L-Shape; Article Numbers: 21482, 21483. 7. FLS - Foot Locking Plates System, FLS Plate, Straight; Article Numbers: 21421-5. 8. FLS - Foot Locking Plates System, Stabilization Screw; Article Numbers: 37241-10, 37241-12, 37241-14, 37241-16, 37241-18, 37241-20, 37241-22, 37241-24, 37241-26, 37241-28, 37241-30. 9. HOL - Hallux Osteotomy Locking Plate, Hallux Osteotomy Plate; Article Numbers: 21015-08, 21015-09.